• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0062101400
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
Note: this report pertains to one of the two nephromax balloon dilatation catheter used during the same procedure.It was reported to boston scientific corporation that two nephromax balloon dilatation catheter devices were used in the kidney during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the balloon was inserted over the wire and was inflated, however; it did not reach 20atm and burst.The same issue occured on the second nephromax balloon dilatation catheter.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Additional Manufacturer Narrative
Block a2: patient's exact age is unknown; however it was reported that the patient was over the age of 18.Block h6: problem code 1074 captures the reportable event of balloon burst.Block h10: investigation results a visual examination of the returned complaint device found that the balloon had a longitudinal tear which started at the distal markerband and extended proximally along the balloon.There was no evidence to suggest that the device was used in a manner inconsistent with the labelled indications.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
Note: this report pertains to one of the two nephromax balloon dilatation catheter used during the same procedure.It was reported to boston scientific corporation that two nephromax balloon dilatation catheter devices were used in the kidney during a percutaneous nephrolithonomy (pcnl) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the balloon was inserted over the wire and was inflated, however; it did not reach 20atm and burst.The same issue occured on the second nephromax balloon dilatation catheter.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPHROMAX
Common Device Name
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8738902
Report Number3005099803-2019-03330
Device Sequence Number19204750
Product Code LJE
UDI-Device Identifier08714729826484
UDI-Public08714729826484
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K121614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/05/2022
Device Model NumberM0062101400
Device Catalogue Number210-140
Device Lot Number0023445286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2019
Initial Date Received by Manufacturer 06/04/2019
Supplement Date Received by Manufacturer07/12/2019
Initial Report FDA Received Date06/27/2019
Supplement Report FDA Received Date08/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-