MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SOLARA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SOLARA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W4TR03
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Erythema (1840); Unspecified Infection (1930); Pocket Erosion (2013); Swelling (2091)
Event Date 06/05/2019
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 429888 lead, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that approximately four months post replacement procedure, the patient developed a pocket erosion.It was noted that the cardiac resynchronization therapy pacemaker (crt-p) eroded through the skin.It was also noted that the skin around the pocket was red and swollen.The crt-p system was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The device was returned and screening analysis was performed, but no issue was identified requiring full analysis.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLARA QUAD CRT-P MRI SURESCAN
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8739371
MDR Text Key149349969
Report Number3004209178-2019-12477
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735682
UDI-Public00643169735682
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/14/2019
Device Model NumberW4TR03
Device Catalogue NumberW4TR03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5086MRI52 LEAD, 5086MRI58 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age80 YR
-
-