|
Batch review performed on 25 june 2019: lot 176627: (b)(4) items manufactured and released on 12 february 2018.Expiration date: 2023-02-01.No anomalies found related to the issue.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs department: acetabular component revision performed 1 year after cementless total hip arthroplasty in a (b)(6) woman.The patient was complaining about pain probably caused by suboptimal cup position.The reason of this position cannot be assessed not having peroperative images: patient anatomy or bone morphology may be a possible cause.No reason to suspect that this event is due to a faulty device.Visual inspection performed by r&d project manager: cup shows ha residuals on outer surface.Signs of extraction on the edge.Pe shows sign of removal.No particular signs on the femoral head.
|