|
Results: the penumbra system 3max reperfusion catheter (3maxc) started to stretch approximately 146.5 cm from the hub.The 3maxc fractured approximately 151.0 cm from the hub.The marker band was present on the fractured distal segment.The total length of the 3maxc including the stretched segment was approximately 164.0 cm.Conclusions: evaluation of the returned 3maxc confirmed a fracture and revealed that the device was stretched.If the device is forcefully retracted against resistance, damage such as stretching may occur.Subsequently, if the extrusion is elongated beyond its stretching length, damage such as a fracture may occur.The jetd identified in the complaint was not returned to penumbra for evaluation, therefore, the root cause of resistance could not be determined.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|