| Model Number |
4FC12 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Cardiac Tamponade (2226); Pericardial Effusion (3271)
|
| Date of Event |
06/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the data files were returned and analyzed.The data files indicated balloon catheter 2af283 with lot number 96237 was used for two applications without issues.The clinical issues (tamponade, effusion, and bradycardia), could not be confirmed through data analysis.Analysis of the physical product is pending upon return of the sheath 4fc12.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cryoablation procedure, the temperature was not as expected.The patient then became irritated and attempted to get off of the table.It was then reported that the patient experienced bradycardia.An echocardiogram was performed identifying an effusion that developed into two tamponades, which were drained prior to the patient being sent to cardiac surgery.It was noted that the patient was in pulseless electrical activity (pea) and that a blood transfusion had to occur.The case was aborted while the patient was under general anesthesia.The patient is expected to make a full recovery.No further patient complications have been reported as a result of this event.
|
| |
|