Catalog Number 5C4476 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was an "eyelash" in an effluent sample bag.This was found during unspecified process step of peritoneal dialysis therapy; however, there was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information : one actual sample was received for evaluation.A visual inspection was performed and it was noted that particulate matter (eyelash) was on the effluent sample bag.During sample evaluation, it was noted that the particulate matter was outside of the effluent sample bag.The reported condition was verified.The direct cause of the event was determined to be manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|