| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Intraocular Pressure Increased (1937); Uveitis (2122); No Code Available (3191)
|
| Date of Event |
05/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight: unk.Ethnicity: unk.Race: unk.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm micl12.6 implantable collamer lens, -12.5 diopter, in the patients left eye (os) on (b)(6) 2019.The lens was explanted on (b)(6) 2019 due to excessive vaulting, pigment dispersion and anterior chamber inflammation.The reporter indicated the excessive vaulting caused pigment dispersion (increasing intraocular pressure controlled on drops and heavy pigment on icl), and anterior chamber inflammation spikes as steroids were tapered.The patient had constant aching sensation but the patient is doing well at one-day post-op explant exam.The cause of the event was due to the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: corrected to (b)(6) 2019 in initial mdr.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: device evaluation: the lens was returned in liquid in a lens vial, in a specimen cup.Visual inspection found a haptic tear.Claim# (b)(4).
|
| |
|