| Model Number |
MMT-551LNAH |
| Medical Device Problem Codes |
Mechanical Problem (1384); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the they experienced hyperglycemia also insulin pump had motor error alarm.The customer blood glucose level was 582 mg/dl.Customer stated they did not feel okay to troubleshoot for high blood glucose.The insulin pump will be returned for analysis.Frn-unk-rsvr, unomed set.
|
| |
|
Additional Manufacturer Narrative
|
|
No motor error alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.(b)(4).
|
| |
|