• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-00
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective motor control board.The technician replaced the board to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event or Problem Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp displayed an error message during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Common Device Name
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key8743008
Report Number9611109-2019-00491
Device Sequence Number19165071
Product Code DWA
UDI-Device Identifier04033817900931
UDI-Public010403381790093111151214
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K032213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60-01-00
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2018
Initial Report FDA Received Date06/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-