• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GERMANY ARCHITECT BNP CONTROLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GERMANY ARCHITECT BNP CONTROLS Back to Search Results
Catalog Number 08K28-12
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction/removal number: 3002809144-06/4/19-006-r.A product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
 
Event or Problem Description
The customer observed biorad controls shift high on the architect bnp assay.The customer indicated the assay was calibrated with control lot 44k82018.There was no report of incorrect patient results or impact to patient management.
 
Additional Manufacturer Narrative
After further evaluation, the suspect medical device architect bnp controls, list 08k28-12, lot 44k82818, manufacturing site (b)(4) is no longer considered a likely cause.Therefore, this medical device report is not related to remedial action 3002809144-06/4/19-006-r and is unassigned from the field action.Refer to mdr 3002809144-2019-00382 which was submitted for the customer's issue on an alternate suspect medical device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT BNP CONTROLS
Common Device Name
BNP CONTROLS
Manufacturer (Section D)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key8743585
Report Number3002809144-2019-00383
Device Sequence Number2296756
Product Code JJX
UDI-Device Identifier00380740145262
UDI-Public00380740145262
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/15/2019
Device Catalogue Number08K28-12
Device Lot Number44K82818
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/07/2019
Supplement Date Received by Manufacturer07/22/2019
Initial Report FDA Received Date06/28/2019
Supplement Report FDA Received Date07/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3002809144-06/4/19-006R
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I1000SR ANALYZER, LN 01L86-40; ARCHITECT I1000SR ANALYZER, LN 01L86-40; SN (B)(4); SN (B)(4)
-
-