| Model Number |
MMT-7020B |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Unexpected Therapeutic Results (1631); Calibration Problem (2890)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Blood Loss (2597)
|
| Date of Event |
06/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
"the customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm." the customer¿s blood glucose was 115 mg/dl and the sensor glucose was 44 mg/dl at the time of the incident.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|