• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020B
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Unexpected Therapeutic Results (1631); Calibration Problem (2890)
Health Effect - Clinical Codes Hypoglycemia (1912); Blood Loss (2597)
Date of Event 06/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
"the customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm." the customer¿s blood glucose was 115 mg/dl and the sensor glucose was 44 mg/dl at the time of the incident.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020B ENLITE3 US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8743627
Report Number2032227-2019-22750
Device Sequence Number1490166
Product Code OZP
UDI-Device Identifier00763000140779
UDI-Public(01)00763000140779(17)200313(10)HG38QJH
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/13/2020
Device Model NumberMMT-7020B
Device Catalogue NumberMMT-7020B
Device Lot NumberHG38QJH
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2019
Initial Report FDA Received Date06/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
Patient Weight132
-
-