• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number MA-15C
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had vibration.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use over time.If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).
 
Event or Problem Description
It was reported from (b)(6) that during evaluation of the bearing sleeve device, it was determined that the device had vibration and was making an unexpected noise.It was noted in the service order that the cutting burr devices were not rotating smoothly.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEARING SLEEVE, 15CM CURVED
Common Device Name
DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key8744028
Report Number1045834-2019-53925
Device Sequence Number13873089
Product Code HBE
UDI-Device Identifier00845384004325
UDI-Public00845384004325
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMA-15C
Device Lot NumberC223021980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2019
Initial Report FDA Received Date06/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-