DEPUY ORTHOPAEDICS INC US UNIVERSAL STEM 75X22MM FLUTED; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS Back to Search Results
Model Number 86-7420
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); No Code Available (3191)
Date of Event 06/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient was revised to address pain.Recently patient was seen in the clinic by the surgeon saying that he heard something snap in his leg.X-ray showed what would appear to be a break at the junction between the femoral sleeve and universal stem.Patient was told to stay off leg and we would plan an additional revision surgery.Patient apparently didn't take advice and went to work next day.A few days later returned to hospital with a painful left leg.X-ray showed a fractured femur at the location of the implant break.The team removed all the femoral implants and revised.Doi: (b)(6) 2019; dor: (b)(6) 2019, left knee.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the device history records were reviewed for product no.867420, lot no.J16j83, description universal stem 75x22mm fluted.And no deviations were noted.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL STEM 75X22MM FLUTED
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8745446
Report Number1818910-2019-97127
Device Sequence Number19342972
Product Code JWH
UDI-Device Identifier10603295217862
UDI-Public10603295217862
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K063633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-7420
Device Catalogue Number867420
Device Lot NumberJ16J83
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2019
Initial Date Received by Manufacturer 06/07/2019
Supplement Date Received by Manufacturer06/24/2019
08/05/2019
02/13/2020
Initial Report FDA Received Date06/28/2019
Supplement Report FDA Received Date07/05/2019
08/07/2019
02/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient Weight186
-
-