Catalog Number 157011165
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 06/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that an unknown white residue was left on the inner packaging after the tyvek material was removed.It was hypothesized that the white reside was excess glue that simply pulled some of the tyvek material off.Either way, there were concerns about the "unknown and a replacement stem was requested.Thank you.2 minutes surgical delay.Doi: unknown.Dor: (b)(6) 2019 left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot; j08y31 device history review: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ9 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8745586
Report Number1818910-2019-97149
Device Sequence Number16476345
Product Code LPH
UDI-Device Identifier10603295060109
UDI-Public10603295060109
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K001991
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number157011165
Device Lot NumberJ08Y31
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/07/2019
Supplement Date Received by Manufacturer07/08/2019
Initial Report FDA Received Date06/28/2019
Supplement Report FDA Received Date07/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-