The device was not returned for analysis.Since the device was not returned, we are unable to perform further root cause analysis.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.No corrective action.Monitoring and trending this type of event.Linked mdrs: 2029214-2019-00644, 2029214-2019-00645, 2029214-2019-00646.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received a report that the coil did not detach from the microcatheter.However, after replacing the coil, the procedure was completed without issue.The anatomy was normal in tortuosity.2 to 3 attempts were made with the instant detacher and 2 to 3 manual detachments were made.There were no issues prior to this event.There was no damage to the pushwire when it was removed.The devices were prepared and used per the ifu.Ancillary devices include a n excelsior 1018 microcatheter.
|
|
D10.Device available, return date - additional information g4.Date manufacturer received - additional information g7.Type of report - additional information h2.Follow-up type - additional information h3.Device evaluation, device returned - additional information h6.Evaluation codes - additional information, device evaluation h10.Additional manufacturer narrative - additional information, device evaluation the concerto coil (model: nv-2-8-helix lot: a697194) was returned for analysis.There was no instant detacher, microcatheter, or accessories returned with the device.The concerto coil was decontaminated.The actuator interface was found bent but still securely attached to the coupler tube.No damages or irregularity was observed with the positive load indicator, coupler tube, break indicator and crimps.There are no indications of any mechanical or manual detachment attempts at these locations.A second bend was found on the pushwire.The implant coil was found damaged but still attached to the pushwire within the introducer sheath.The introducer sheath was found correctly assembled on the pushwire with the wavelock positioned on the proximal end.No damage was found with the introducer sheath.The coin was found to be present and pushed up against the lumen stop; with the shield coil present and intact.Mechanical detachment with an in-house instant detacher was not performed due to the damaged actuator interface.A manual detachment was attempted by breaking the break indicator and pulling back the release wire and the implant coil was successfully detached with no issue.Based on the analysis performed, the concerto coil was confirmed to have non-detachment as the pushwire was returned with the implant coil still attached, however the cause cannot be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|