| Model Number |
401763 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was chosen to treat a severely calcified in an lad.During procedure the stent dislodged.
|
| |
|
Additional Manufacturer Narrative
|
|
A correction was made to the event description.This stent did not dislodge from the balloon.The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the orsiro stent is neither deformed nor damaged.The stent is properly crimped in between the two radiopaque markers.The balloon is well folded and shows no signs of inflation.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The final root cause is most likely related to external factors.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was chosen to treat a severely calcified lesion in an lad but the target lesion could not be crossed.
|
| |
|