| Model Number |
GF-210RA |
| Medical Device Problem Codes |
Loss of Data (2903); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer would like to send the unit in for repair.The device was in use on a patient, but no patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the multigas unit was dropping patient data intermittently during cases.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the multigas unit was dropping patient data intermittently during cases.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: on (b)(6) 2019 customer stated gas unit dropped out intermittently.Nk repair center evaluated the device and noted device was giving "gas device error" on the monitoring device.Service requested: exchange.Service performed: exchange.Investigation result: the investigation under irc-nka300155492 concluded that for gf-210ra revision cb and onward with sensor unit cd-314p revision 2 (serial numbers (b)(6)) needed a firmware upgrade.Revision 2 of cd-314p utilize a new pump with a slight difference in specification.The sensor unit detects this small difference and output as gas device error.The resolution was to perform a firmware upgrade.For gf-210ra with older sensor unit cd-314p (no revision # = first revision), there was no tendency of sensor failing, thus the investigation provided no countermeasure.Due to no tendency of failure, it was determined these sensors failed as a result of normal usage.Cd-314p replacement is recommended.In summary, serial numbers(b)(6)and below has the first version of cd-314p.Serial numbers (b)(6) have version 2 of cd-314p.These units require firmware upgrade.Serial numbers above (b)(6) are not affected.The root cause of the error message on serial number 1262 was due to firmware requiring upgrade.
|
| |
|
Search Alerts/Recalls
|