The electronic log file indicates that the device passed the power-on self-test in the morning of the date of event w/o deviations.The particular procedure suffered right from the beginning from instable pressure conditions and fluctuating tidal volumes.Several alarms like pinsp not attained, apnea and fresh gas low / leakage were posted.After 8 minutes into the procedure the software detected a stalling of the ventilator piston due to a pressure peak in the range of 65 mbar.Consequently, a shutdown of automatic ventilation was forced which was brought to the user's attention by means of a corresponding alarm.Patient support was continued in manual ventilation for another nine minutes before the device was set to standby.No indications for a device malfunction was found in the logs.The unit was tested in follow-up of the event by a technician; no problems were detected either during the testing and the consecutive endurance run.Dräger finally concludes that the stop of ventilation was not related to a malfunction but furthermore the system's response to an external disturbance to protect from potential risks to the patient and to avoid damages of the ventilation unit.The exact root cause remains unknown; however - imaginable scenarios would be patient coughing, spontaneous breathing or chest movements caused by patient repositioning etc.There's no issue with the device that would require repair or correction.
|