| Model Number |
MMT-723NAL |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump received compromised force sensor system.The customer¿s blood glucose level was unknown.Troubleshooting was not performed.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/flush drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test due to motor error alarm.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported via phone call that the weight has been changed to 0156.
|
| |
|