On 06/25/2019, a distributor of infutronix reported a tubing bonding issue.The distributor received the following information from the user facility on 06/25/2019 that over the night the patient "got up in the middle of the night, dropped the pump and the tubing on the outflow side of the pump came out of the pump.He called customer service after reinserting the tubing and was told to turn off the pump, clamp the tubing and come in to see us first thing this morning.Other than losing 5 hours of infusion time, there was minimal leakage and no other problems.We replaced the tubing this morning and sent him home with the infusion resumed." the distributor reported that no patient was harmed and the medication infused was 5fu.The user facility representative reported that the lot number of the affected device was probably 1901003.The contract manufacturer of the affected device is (b)(4).
|