DEPUY SYNTHES PRODUCTS LLC 8 CM HEAVY DUTY MEDIUM ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number MEDIUM-HD-G1
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).The device manufacture date was unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event or Problem Description
It was reported from (b)(6) that the inner metal sheath of the attachment device fell out.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device manufacture date was documented as unknown in the initial report.The device manufacture date has been updated as 5/2/2018.Device evaluation: the actual device was returned for evaluation.A visual and functional assessment was performed on the device which determined that the inner components were missing.It was determined that the issue was consistent with overtime wear.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear out from use.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8 CM HEAVY DUTY MEDIUM ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key8748212
Report Number1045834-2019-53937
Device Sequence Number14052923
Product Code HBC
UDI-Device Identifier00845384016328
UDI-Public00845384016328
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMEDIUM-HD-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2019
Initial Date Received by Manufacturer 06/21/2019
Supplement Date Received by Manufacturer08/05/2019
Initial Report FDA Received Date07/01/2019
Supplement Report FDA Received Date08/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-