| Catalog Number |
08K28-11 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Correction/removal number: 3002809144-06/4/19-006-r a product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
|
| |
|
Event or Problem Description
|
|
The customer observed abbott controls shift high and out of range while using the architect bnp control lot 44k80918.There was no report of incorrect patient results or impact to patient management.
|
| |
|
Search Alerts/Recalls
|