Film analysis summary: the exact cause of the reported events could not be determined from the two (2) still screenshot angiograms returned.The anatomy prior to the 2008 talent aaa implant is unknown, and follow-up films during the implant duration were not available.Ct¿s just prior to the recent navion intervention were also not available for review; therefore, no assessment of the implanted talent aaa stent graft system could be performed.It is likely that the reported disease progression, continued aortic elongation, led the need to extend the device proximally using the navion/taa devices.The reported tip detachment occurring during removal of the delivery system was confirmed.However, the exact cause of the removal difficulties, which appeared to have been due to the tip becoming caught on the talent contralateral limb, could not be determined from the limited returned films.It appears likely that this event was due to a user issue (not observing during removal through the previously implanted aaa limb, that may have been exacerbated by an anatomical issue.The returned angiogram did not show the talent devices prior to implanting the navion, and images during the navion implant and removal of the delivery system (when the tip became caught inside the talent limb) were not available for review.It is possible that the observed contra limb separation from the bifurcate stub, near to where the detached tip was located, also occurred during the delivery system removal difficulties.A device issue cannot be ruled out as a potential cause.However, the delivery system was discarded and a product investigation could not be performed.The cause of the observed occlusion seen within the detached contra limb, and the reason for the reported fem-fem bypass also could not be determined from the returned images.If information is provided in the future, a supplemental report will be issued.
|
|
A valiant navion stent graft system was implanted in a patient for the endovascular treatment of a 40mm diameter thoracoabdominal aneurysm.The patient had a previous talent abdominal stent graft system implanted approximately eleven years earlier.It was noted that there was no failure to the previous graft and the aorta had continued to elongate.It was reported that during the index procedure, the physician placed the device and attempted to remove the delivery system, going from the grey handle to the blue, however it would not advance.The physician then twisted the device without looking distally through the previously implanted aaa device.It was then noted that the aaa limb had wrapped around the navion delivery system.The delivery system could not be removed from the patient.The physician then broke down the delivery system to attempt troubleshooting, however the delivery system could still not be removed from the patient.At this point, the physician performed a cut down on the access site on the patient's right side.They then cut the delivery system, leaving the nose cone and tip capture inside the patient.A fem-fem bypass was then carried out.Ten days later further intervention was carried out.The physician was able to deploy the last two stent grafts successfully and seal was obtained resolving a type ia endoleak.No attempt to remove the nose cone was completed during this procedure.As per the physician, the cause of the event could not be determined.No additional clinical sequelae were reported and the patient is fine.
|