| Model Number |
799-EGL2-00 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including full functionality testing without duplicating the report.The autocal valve was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "runtime calibration failure - 1051" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|