| Model Number |
694765 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Fracture (1260); High impedance (1291); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
06/13/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited high impedance and a sensing problem due to a lead fracture.The patient received inappropriate therapy as a result.The lead was capped and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the proximal portion of the lead was returned, analyzed, and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|