| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the reporter contacted animas, alleging a call service alarm (cs 064) issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the pump has been returned and evaluated by product analysis on 01-aug -2019 with the following findings: a review of the pump¿s black box showed evidence of a cs 064 motor alarm on (b)(6)2019.During testing, the pump successfully completed the rewind, load cartridge, and prime steps and the pump was exercised for 12 hours with no alarms occurring.The pump cover was removed and moisture corrosion was found on the motor flex connector.The cs 064 call service alarm issue was shown in the pump history but was not duplicated during investigation.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|