• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZYNO MEDICAL, LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZYNO MEDICAL, LLC ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number B2-70071-DF
Medical Device Problem Codes Fluid/Blood Leak (1250); Leak/Splash (1354)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Information (3190)
Date of Event 05/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The facility representative will send in an unused administration set that he believed to be associated with the leaking event for evaluation.The unused administration set has a model number of b2-70071-df, a lot number of 18045836, and an expiration date of 04/13/2021.
 
Event or Problem Description
Zyno medical received a filter leaking issue reported from a distributor on 06/24/2019.On (b)(6) 2019, the distributor received the initial report from a user facility representative.The user facility representative stated that the leaking was from the 0.2 micron filter on the b2-70071-df administration set.He also stated the leaks resulted in the loss of tens of thousands of dollars of medication, and some of the medication was toxic.The defective administration set had been discarded by the user facility.The associated complaint number is (b)(4) in zyno's record.The event date was reported as (b)(6) 2019.It was unknown if there was patient involvement.No patient harm was reported and the patient information was unavailable.The medication used at the time of event was unknown but it was a type of chemo.No lot number of the affected device was provided, and no expiration date or manufacturing date was available.The distributor had requested the user facility to obtain the lot number information from the nurses if possible and save any future leaky administration sets in biohazard bags for returns and evaluation.The contract manufacturer of the affected device is becton, dickinson and company.
 
Additional Manufacturer Narrative
The contract manufacturer provided the evaluation report to zyno medical on 03/10/2020.A trend review was performed.5 additional complaints related to the failure mode were found in a 1-year period; 2 quality notifications related to this failure mode were found in a 1-year period.However, due to the fact that the actual affected device was discarded by the user, the investigation couldn't be completed.The root cause couldn't be confirmed.No preventive or corrective action was taken.The contract manufacture also performed the investigation on the customer returned unused administration set sample, which the customer believed to be associated with the leaking event.The sample has a model number of b2-70071-df and a lot number of 18045836.The failure mode couldn't be repeated.No issue was identified on the sample.The sample device performed as expected.The complaint couldn't be confirmed.
 
Event or Problem Description
This is a follow-up for the initially filed mdr 3006575795-2019-00015.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL ADMINISTRATION SET FOR Z-800 INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
ZYNO MEDICAL, LLC
177 pine street
natick MA 01760
MDR Report Key8750214
Report Number3006575795-2019-00015
Device Sequence Number2338862
Product Code FPA
UDI-Device Identifier00814371020167
UDI-Public00814371020167
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K132841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB2-70071-DF
Device Catalogue NumberB2-70071-DF
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2019
Supplement Date Received by Manufacturer03/10/2020
Initial Report FDA Received Date07/01/2019
Supplement Report FDA Received Date03/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-