MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 19MM MOD REV HIP BDY/BLT +10MMCOMPONENT LEVEL 9006-1-119; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 19MM MOD REV HIP BDY/BLT +10MMCOMPONENT LEVEL 9006-1-119; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6276-1-119
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 04/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Rep reported through a fax that patient dislocated their right hip.Open reduction of hip, changed adm liner and femoral head.Update 04 june, 2019: operative report states "it was found that the proximal body was placed in significant anteversion, so then if the patient came in extension it would lever out anteriorly".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
19MM MOD REV HIP BDY/BLT +10MMCOMPONENT LEVEL 9006-1-119
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8750320
Report Number0002249697-2019-02446
Device Sequence Number1708793
Product Code LPH
UDI-Device Identifier04546540641137
UDI-Public04546540641137
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number6276-1-119
Device Lot Number62514102
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/04/2019
Initial Report FDA Received Date07/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
Patient Weight141
-
-