| Catalog Number |
6276-1-119 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Codes |
Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
04/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Rep reported through a fax that patient dislocated their right hip.Open reduction of hip, changed adm liner and femoral head.Update 04 june, 2019: operative report states "it was found that the proximal body was placed in significant anteversion, so then if the patient came in extension it would lever out anteriorly".
|
| |
|