| Catalog Number |
1235-2-502 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
06/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
As reported: "left tha revision due to dislocation.Patient experienced dislocation.During the attempt to reduce her hip, the force being applied pried the adm insert off of the 28mm head".Spoke to rep.The restoration adm insert disassociated out of the restoration adm shell.Patient went to er where a closed reduction was attempted.During the reduction the head dislocated out of the liner and the patient was revised.
|
| |
|