As reported, during the operation when filling contrast agent, leakage was noted from the saber (3 mm15 cm 150) balloon catheter.The user immediately stopped using the saber and it was replaced with a new device to complete the operation.There was no reported patient injury.The product was clinically used and will be returned for analysis.
|
|
Complaint conclusion: as reported, during the operation when filling contrast agent, leakage was noted from the saber (3mm x 15cm x 150cm) balloon catheter.The user immediately stopped using the saber and it was replaced with a new device to complete the operation.There was no patient injury.The product was not used inside the patient.The intended procedure was a lower limb artery angiographic stent implantation.The lesion was calcified, with no tortuosity, and sixty percent stenosed.The device was meant to be used for a chronic total occlusion (total occlusion >3 months).The leakage was noted during prep and leaking from the balloon.There were no kinks or other anomalies noted when the device was taken out of the package.The device was not re-sterilized.The device was prepped normally (i.E.Maintained negative pressure) following ifu instructions.There was no difficulty removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.Saline contrast media was used at a 1:1 ratio.The same indeflator was used successfully with other devices.The catheter was never in an acute bend.The product was returned for analysis.A non-sterile saber 3mm x 15cm x 150cm pta balloon catheter was returned.Per visual analysis, the balloon catheter was coiled inside a plastic bag and was received deflated.The body shaft was bent at 51.5cm and at 56.3cm from the strain relief.No other anomalies were observed.Functional test analysis was performed, a three-way valve was attached to the inflation lumen port.Required pressure was applied to the device to inflate the balloon, no issues were observed.A product history record (phr) review of lot 17714713 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon - leakage¿ was not confirmed by analysis of the returned device.The exact cause of the reported event could not be confirmed.The device performed as intended as functional analysis was performed successfully.No anomalies were observed on the balloon.Based on the information available for review, it is difficult to draw a clinical conclusion between the device and the reported event.However, procedural factors and handling of the device may have contributed to the reported event.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
|