MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, DIRECT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, DIRECT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0279350401
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the tip broke during the procedure.All pieces were retrieved and the procedure was completed successfully.
 
Event or Problem Description
It was reported that the tip broke during the procedure.All pieces were retrieved and the procedure was completed successfully.
 
Additional Manufacturer Narrative
Alleged failure: the tip of the electrode was broken off the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting of removing probe, probe inserted into obstructed passageway or any forceful impact to the tip of the probe with rigid objects.Probe used as a tool for mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES, DIRECT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8751160
Report Number0002936485-2019-00266
Device Sequence Number1690516
Product Code GEI
UDI-Device Identifier07613327056921
UDI-Public07613327056921
Combination Product (Y/N)N
PMA/510(K) Number
K071859
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279350401
Device Lot Number18086AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2019
Initial Date Received by Manufacturer 06/04/2019
Supplement Date Received by Manufacturer06/04/2019
Initial Report FDA Received Date07/01/2019
Supplement Report FDA Received Date08/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-