• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 220; JCX Back to Search Results
Catalog Number PMX220
Medical Device Problem Codes Misconnection (1399); Device Handling Problem (3265)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system aspiration pump max 220 (pump max).During the procedure, the hospital staff accidentally connected the aspiration tubing (tubing) directly to the pump max rather than the canister.Consequently, blood was aspirated into the pump max and it stopped working; therefore, the pump max was disconnected.The procedure was completed using another pump max.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: a pipe cleaner was inserted into the vacuum inlet and blood was found inside the penumbra system aspiration pump max 220 (pump max) assembly.Conclusions: evaluation of the returned pump max confirmed fluid within the pump assembly.If the tubing is connected directly to the vacuum inlet, blood may be aspirated into the pump max.If fluid is aspirated into the pump max, the device may not function properly.Penumbra pumps are 100% visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 220
Common Device Name
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8753797
Report Number3005168196-2019-01310
Device Sequence Number2037964
Product Code JCX
UDI-Device Identifier00814548012698
UDI-Public00814548012698
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K122756
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPMX220
Device Lot NumberF13287-17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2019
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/06/2019
Supplement Date Received by Manufacturer07/18/2019
Initial Report FDA Received Date07/02/2019
Supplement Report FDA Received Date08/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-