| Catalog Number |
TRX-340 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation due to capsular contracture repair, baker grade unknown.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side right breast tenderness, bloody right nipple discharge and capsular contracture baker grade unknown.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Search Alerts/Recalls
|