• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. COVATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. COVATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number AC4340-01
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during an unknown procedure, the black end of the wands cracked.The procedure was completed with a backup device with no delay or patient injury reported.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation.A relationship, if any, between the subject devices and the reported event could not be determined since the products were not returned for evaluation.Visual inspection and functional testing could not be performed because the devices in question were not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the products associated with this event are returned at a future date, this evaluation will be reopened for investigation.Factors not associated with the manufacture or design of the devices which could result in damage to the tips could be the devices coming into abrupt contact with a hard (metallic) object, improper insertion or removal from an apparatus, or excessive force applied to the tips can also lead to unintended detachment.There were no indications during manufacturing record review that would suggest that the devices did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVATOR
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8755811
Report Number3006524618-2019-00314
Device Sequence Number19259436
Product Code GEI
UDI-Device Identifier00817470003314
UDI-Public00817470003314
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2021
Device Catalogue NumberAC4340-01
Device Lot Number2023449
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/12/2019
Supplement Date Received by Manufacturer07/31/2019
Initial Report FDA Received Date07/02/2019
Supplement Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-