• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIOSURE HA 10MM X 25MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIOSURE HA 10MM X 25MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 72201785
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during a reconstruction of the anterior cruciate ligament (right knee), the thread of the screw broke during the implantation, the pieces were removed using tweezers.There was a surgical delay greater than 30 minutes.A backup device was available to complete the procedure with no patient injuries.
 
Additional Manufacturer Narrative
One 10x25 biosure-ha screw was returned for evaluation.Visual assessment of the screw confirmed the reported complaint of breakage.Approximately.5mm of the distal threads have broken off and were not returned for examination.Clinical details regarding site preparation, graft type and patient bone quality were not supplied.The condition of the device indicates the screw was subjected to excessive force.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: improper site preparation, incorrect screw size.The instruction for use was reviewed and was found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 10MM X 25MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8756125
Report Number1219602-2019-00772
Device Sequence Number2280762
Product Code HWC
UDI-Device Identifier03596010611819
UDI-Public03596010611819
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2023
Device Catalogue Number72201785
Device Lot Number50774354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Initial Date Received by Manufacturer 06/13/2019
Supplement Date Received by Manufacturer07/08/2019
Initial Report FDA Received Date07/02/2019
Supplement Report FDA Received Date07/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age33 YR
Patient Weight82
-
-