COOK ENDOSCOPY TRI-TOME PC PROTECTOR; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
| Model Number |
G49197 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: non-healthcare professional.Investigation evaluation: an evaluation of the returned device confirmed the report based on a breakage observed at the center cannula, but could not confirm the report on incorrect orientation.The breakage at the center cannula prevented bowing of the device.During our laboratory analysis, the sphincterotome was advanced through a duodenoscope that was placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter (olympus model number tjf-160v).The catheter exited the endoscope with the cutting wire facing 12 o¿clock.The device was then attempted to be bowed but the distal end would not respond to handle manipulation.The distal end would not respond to handle manipulation due to a breakage at the center cannula joint.The sphincterotome catheter was subjected to a close visual examination and twisting of the tubing was not observed at the distal end.A discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Prior to distribution, all tri-tome pc protector sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook tri-tome pc protector sphincterotome.The user found out the orientation of the cutting wire was not centered during cutting [of the] duodenal papilla, and may cause cutting position not [to be]accurate [incorrect cutting wire orientation].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|