• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R FEM LT SZ 4 TRL; KNEE INSTRUMENT : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R FEM LT SZ 4 TRL; KNEE INSTRUMENT : FEMORAL TRIALS Back to Search Results
Catalog Number 961004
Medical Device Problem Codes Device-Device Incompatibility (2919); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the size 4 sigma notch trial is stuck in the size 4 (left) sigma femoral component.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.The noted damage is consistent with the use of excessive torque and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8757642
Report Number1818910-2019-97603
Device Sequence Number1680872
Product Code JWH
UDI-Device Identifier10603295234586
UDI-Public10603295234586
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K943462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number961004
Device Lot NumberJ0605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2019
Initial Date Received by Manufacturer 06/17/2019
Supplement Date Received by Manufacturer08/15/2019
Initial Report FDA Received Date07/03/2019
Supplement Report FDA Received Date08/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-