• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE; MAGIC 3 HYDROPHILIC INTERMITTENT CATHETER WITH GRIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE; MAGIC 3 HYDROPHILIC INTERMITTENT CATHETER WITH GRIP Back to Search Results
Model Number 53618G
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Urinary Tract Infection (2120); Patient Problem/Medical Problem (2688)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the magic 3 hydro catheters were stiff and sharp rendering them hard to insert.The patient experienced self-limited bleeding.The patient was able to void with catheters, however, he allegedly contracted an urinary tract infection from use of the catheter.The patient was treated with antibiotics.
 
Additional Manufacturer Narrative
The reported event was unconfirmed.Visual inspection noted 4 unopened magic 3 intermittent catheters were received.Visual evaluation noted no sharp points or flash.The catheters felt normal.Further testing required.The samples were tested according to the general inspection level ii/single sampling plan.The catheters' od measured to and were found to be within specification (0.236" +/- 0.013").The catheters' id measured and were found to be within specification (0.147" +/- 0.013").The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following:¿instructions for use intended use: the catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/ precautions and adverse reactions.Wash your hands thoroughly with soap and water.Release the sterile water from the foil packet.Tip the catheter pouch end-to-end three to six times so the water moves back and forth to thoroughly wet the catheter surface.Peel open the pack at the funnel end just enough to expose the insertion sleeve.Don¿t remove the catheter yet.Use the adhesive tab at the funnel end of the pack to stick the pack to a nearby vertical surface while preparing to catheterize.Wash the area around the meatus before catheterizing.Wash your hands again.Hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Repeat until urine starts to flow.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.Finish by disposing of the catheter and its packaging.Wash your hands with soap and water.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Urethral catheter for urological use only.Discard after use.Made of silicone elastomer." additional patient code(s): (b)(4).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the magic 3 hydro catheters were stiff and sharp rendering them hard to insert.The patient experienced self-limited bleeding.The patient was able to void with catheters, however, he allegedly contracted an urinary tract infection from use of the catheter.The patient was treated with antibiotics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE
Common Device Name
MAGIC 3 HYDROPHILIC INTERMITTENT CATHETER WITH GRIP
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8757897
Report Number1018233-2019-03592
Device Sequence Number1342192
Product Code KOD
UDI-Device Identifier00801741069796
UDI-Public(01)00801741069796
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/28/2023
Device Model Number53618G
Device Catalogue Number53618G
Device Lot NumberJUCZ0047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2019
Initial Date Received by Manufacturer 06/13/2019
Supplement Date Received by Manufacturer07/16/2019
Initial Report FDA Received Date07/03/2019
Supplement Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-