| Model Number |
53618G |
| Medical Device Problem Code |
Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Urinary Tract Infection (2120); Patient Problem/Medical Problem (2688)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the magic 3 hydro catheters were stiff and sharp rendering them hard to insert.The patient experienced self-limited bleeding.The patient was able to void with catheters, however, he allegedly contracted an urinary tract infection from use of the catheter.The patient was treated with antibiotics.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual inspection noted 4 unopened magic 3 intermittent catheters were received.Visual evaluation noted no sharp points or flash.The catheters felt normal.Further testing required.The samples were tested according to the general inspection level ii/single sampling plan.The catheters' od measured to and were found to be within specification (0.236" +/- 0.013").The catheters' id measured and were found to be within specification (0.147" +/- 0.013").The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following:¿instructions for use intended use: the catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/ precautions and adverse reactions.Wash your hands thoroughly with soap and water.Release the sterile water from the foil packet.Tip the catheter pouch end-to-end three to six times so the water moves back and forth to thoroughly wet the catheter surface.Peel open the pack at the funnel end just enough to expose the insertion sleeve.Don¿t remove the catheter yet.Use the adhesive tab at the funnel end of the pack to stick the pack to a nearby vertical surface while preparing to catheterize.Wash the area around the meatus before catheterizing.Wash your hands again.Hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Repeat until urine starts to flow.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.Finish by disposing of the catheter and its packaging.Wash your hands with soap and water.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Urethral catheter for urological use only.Discard after use.Made of silicone elastomer." additional patient code(s): (b)(4).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the magic 3 hydro catheters were stiff and sharp rendering them hard to insert.The patient experienced self-limited bleeding.The patient was able to void with catheters, however, he allegedly contracted an urinary tract infection from use of the catheter.The patient was treated with antibiotics.
|
| |
|