| Model Number |
PROPAQ MD |
| Medical Device Problem Code |
Output above Specifications (1432)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing which included bench handling, power cycling, full defib functionality testing without duplicating the malfunction.The device was recertified and returned to the customer.No trend is associated with reports of this type.Review of the device activity logs did not show any faults that could be associated with customer report.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device's defib output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|