• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP497G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device mg2958, and no non-conformance's were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that the patient underwent an unknown orthopedic surgery on an unknown date and suture was used.During the procedure, the suture was broken during use.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 08/14/2019.Additional h3; it was reported performance breakage suture.An unopened sample was returned for evaluation.The complaint sample was not received for analysis.During the visual inspection of the sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative sample, no performance - breakage suture was found and the tested sample met the finished goods requirements.Representative samples returned to support investigation of 2 device issues captured in reports 2210968-2019-83640 and 2210968-2019-83641.Analysis results have been included in follow-up reports for each.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 18IN 3-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8758300
Report Number2210968-2019-83640
Device Sequence Number19204676
Product Code GAM
UDI-Device Identifier10705031053816
UDI-Public10705031053816
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Catalogue NumberVCP497G
Device Lot NumberMG2958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Initial Date Received by Manufacturer 06/11/2019
Supplement Date Received by Manufacturer07/22/2019
Initial Report FDA Received Date07/03/2019
Supplement Report FDA Received Date08/14/2019
Patient Sequence Number1
-
-