| Catalog Number |
AB35W06060135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician was using an evercross pta catheter with a trailblazer support catheter to treat a severely calcified, moderately tortuous lesion in the mid proximal common iliac artery.There was no damage noted to the packaging and there were no issues noted when removing the devices from their tray/hoop.The devices were prepped per ifu without issue.No embolic protection was used.The evercross was inflated to 8atm with a non-medtronic inflation device.The device did not pass through a previously deployed stent, but resistance was encountered but no excessive force used on advancing the device to the target lesion.The evercross is reported to have burst.The 2 devices were removed together and then it was noted that the trailblazer delivery catheter was damaged.All fragments of the burst balloon were retrieved.The physician decided to complete the procedure at a later date.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the evercross balloon and trailblazer support catheter were returned for evaluation.A 6f introducer sheath was included the returned contents: no other ancillary devices were included.The evercross catheter was loaded inside the 6f introducer sheath.The sheath was inspected and observed the distal rim was stretched outward.The stretched rim was consistent with encountered ancillary device interaction.The evercross showed a radial balloon burst.Buckling of the catheter was noted within the balloon segment area.The distal end of the balloon material was compressed towards the distal end.The marker band was discovered approximately 3 cm from the tip and showed buckling of the catheter shaft at both edges of the marker band.The proximal segment of the balloon showed creases throughout, but no signs of buckling.The proximal marker band was identified approximately 7cm from the distal tip.The working length of the evercross catheter was approximately 138cm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: minimal resistance was experienced during advancement of the evercross.Severe resistance was encountered during removal.The structural integrity of the trailblazer was reported to be compromised internally.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|