| Model Number |
403087 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An orthalign distributor reported that during a total hip arthroplasty there was a malfunction with the reference sensor.An investigation will be preformed once the reference sensor is returned from orthalign inc.'s distributor in (b)(4).With an abundance of caution this report is being filed with the understanding of the potential patient harm that could result from an accuracy malfunction.A follow up report will be filed with the investigation findings.
|
| |
|
Event or Problem Description
|
|
It was reported that total hip arthroplasty was performed with hipalign.Surgeon installed a cup to the patient hip with impactor connected to the reference sensor.However, angle of the cup was shown abnormally after the cup installation.Subsequently, conventional method was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
No fault could be found with the returned reference sensor, rs s/n: (b)(6).The reference sensor passed all visual and functional tests.A review of the device history record (dhr) was conducted.The device passed all manufacturing specifications prior to release.Orthalign, inc.Will continue to monitor this issue and take action if alert limits are exceeded.
|
| |
|