• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACTIGRAPH, LLC ACTIGRAPH GT9X LINK; EXERCISER, MEASURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACTIGRAPH, LLC ACTIGRAPH GT9X LINK; EXERCISER, MEASURING Back to Search Results
Model Number GT9X LINK
Health Effect - Clinical Codes Reaction (2414); Partial thickness (Second Degree) Burn (2694)
Date of Event 06/05/2019
Event or Problem Description
Actigraph received info from a customer that a participant experienced a skin reaction on "burn" under the metal electrical contacts of a device.This device was not worn in the normal fashion, the wrist holster, but instead physically secured to the user's chest via a tegaderm patch for 48 hours.The display of the device was facing out and the device contact pins were directly touching the skin.The customer reported, the electrodes [on the back of the device] were clearly corroded and there was a greenish patina on / in the participant's skin.The participant also reported the "device felt warm (even indoors in ac)." however, the reporter clarified that it was not a heat related injury, but speculated it appeared to be more of a chemical and / or electrical reaction with metal contacts; specifically the leftmost electrode.The participant reported they sought medical treatment and the physician described the reaction as a "2nd degree burn" and prescribed an antibiotic topical ointment.It was reported to actigraph that a second participant within the same study notified them of having similar sores but was much less severe and they did not seek medical attention; it should be noted this participant was using the device in the same manner.No further pt complications have been reported as a result of this event.The instructions for use of the actigraph gt9x link informs the user to insert the device into the holster and then attach to the wrist or waist during use.In this event, the gt9x link was not put into a holster, but instead physically secured to the user's chest via a tegaderm patch for 48 hours.We advised the reporting organization to not deploy the device in this fashion.This device has been in the field for years with no similar reports.Actigraph did not review any medical records.The physical devices have been returned to actigraph.There is no way to determine the exact cause of skin reactions by analysis of the physical, but a physical product analysis and independent chemical analysis will be performed to investigate any potential chemical which may cause corrosion.The independent chemical analysis is currently ongoing and a supplemental report will be issued if necessary.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIGRAPH GT9X LINK
Common Device Name
EXERCISER, MEASURING
Manufacturer (Section D)
ACTIGRAPH, LLC
pensacola FL 32502
MDR Report Key8760345
Report NumberMW5087861
Device Sequence Number16737815
Product Code ISD
UDI-Device Identifier00853048008015
UDI-Public(01)00853048008015
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Voluntary
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberGT9X LINK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
Patient Weight66
-
-