• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 106A3
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files indicated the electrophysiology (ep) catheter 227f3 with lot number 872218 was used for one application and system notice 50005, indicating that the safety system detected fluid in the catheter and stopped the injection, was received.In conclusion, the reported system notice 50005, indicating that the safety system detected fluid in the catheter and stopped the injection, is not likely to cause or contribute to a death or serious injury; however, the risk of the patient being under general anesthesia without full therapeutic effect is the adverse event being reported.The decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.The coaxial umbilical cable, electrical umbilical cable and electrophysiology (ep) catheter were replaced, and the console was rebooted, without resolve.Another system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.The same products were replaced without resolve.The case was aborted with the patient under general anesthesia.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8760423
Report Number3002648230-2019-00474
Device Sequence Number2040191
Product Code LPB
UDI-Device Identifier00643169449596
UDI-Public00643169449596
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P020045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2019
Initial Report FDA Received Date07/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-