• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-FREIBURG LARYNGEAL ELECTRODE SELECT UNIV CABLE; SURGICAL NERVE STIMULATOR/LOCATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-FREIBURG LARYNGEAL ELECTRODE SELECT UNIV CABLE; SURGICAL NERVE STIMULATOR/LOCATOR Back to Search Results
Model Number 5140-530-866
Medical Device Problem Codes Fracture (1260); Loose or Intermittent Connection (1371); Connection Problem (2900)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the device was damaged and could not provide accurate connection during a surgical procedure at the user facility.This condition, in which the electrical circuit is incomplete, may result in the loss of nerve monitoring recording function, posing increased risk of transection of a nerve.The procedure was completed successfully.There was no surgical delay, no medical intervention, and no adverse consequences.
 
Event or Problem Description
It was reported the device was damaged and could not provide accurate connection during a surgical procedure at the user facility.This condition, in which the electrical circuit is incomplete, may result in the loss of nerve monitoring recording function, posing increased risk of transection of a nerve.The procedure was completed successfully.There was no surgical delay, no medical intervention, and no adverse consequences.
 
Additional Manufacturer Narrative
Additional information: h4-manufacture date follow-up report submitted to document device evaluation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARYNGEAL ELECTRODE SELECT UNIV CABLE
Common Device Name
SURGICAL NERVE STIMULATOR/LOCATOR
Manufacturer (Section D)
STRYKER INSTRUMENTS-FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
MDR Report Key8760478
Report Number0001811755-2019-02209
Device Sequence Number1541572
Product Code ETN
UDI-Device Identifier07613327298086
UDI-Public07613327298086
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K091874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5140-530-866
Device Catalogue Number5140-530-866
Device Lot Number18F013R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2020
Initial Date Received by Manufacturer 06/10/2019
Supplement Date Received by Manufacturer07/28/2020
Initial Report FDA Received Date07/03/2019
Supplement Report FDA Received Date12/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-