• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770500
Device Problems Temperature Problem (3022); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per office manager, machine freezes.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the unit freezes is unconfirmed.The scanner ran for 3 hours and cycled power multiple times.Scanner never stopped functioning as expected.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per office manager, machine freezes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8760690
MDR Text Key150137825
Report Number3006260740-2019-01875
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741096891
UDI-Public(01)00801741096891
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770500
Device Catalogue Number9770500
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2019
Event Location Hospital
Date Manufacturer Received09/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-