| Model Number |
439688 |
| Medical Device Problem Codes |
Failure to Capture (1081); Capturing Problem (2891); Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266)
|
| Health Effect - Clinical Codes |
Muscle Stimulation (1412); Cardiomyopathy (1764); Dyspnea (1816); Edema (1820); Unspecified Infection (1930); Pulmonary Edema (2020); Swelling (2091); Heart Failure (2206)
|
| Date of Event |
05/15/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was experiencing diaphragmatic stimulation.The left ventricular (lv) lead was explanted and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information reported the patient was experiencing worsening heart failure, cardiomyopathy, swelling, edema, shortness of breath, was being treated for a respiratory infection suspected to be related to fluid accumulation, and a chest x-ray noted pulmonary edema.The patient was treated with an increase in oral diuretics.It was further reported that the diaphragmatic stimulation was positional and reprogramming was performed prior to explant.The lv lead also exhibited loss of capture, a rise in thresholds, and high thresholds.The patient is a participant in the post approval clinical surveillance product surveillance registry.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the proximal portion of the lead was returned, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the lv lead was capped, and placement of two new competitor epicardial lv leads was performed.Approximately sixteen months later, during a competitor leads revision procedure, at the site of the capped lv lead, an old full infection was noted.Given the signs of full infection, antibiotic medication was administered.The pocket was irrigated with saline solution, and the cap cleaned and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5.Desc evt problem.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|