| Model Number |
3650-100 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On completion of the investigation a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Emergency room used a oasis ats 3650-100 drain and says that the clinical team complained of lack of flow/drainage with the system.They did not replace the chest drain and continued to use it with no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
As the oasis chest drain was not returned atrium medical corporation cannot conduct a proper investigation of the physical drain.Without the actual drain in question the complaint cannot be confirmed.All chest drains are 100% pressure tested to ensure there are no system leaks prior to packaging.Based on the details of the complaint it is not know what settings the vacuum source was set to.The suction source should be set to -80 mmhg or higher for chest drain regulator settings of -20 cmh2o or greater.This cannot be confirmed without the actual drain that was used.Based on the results of the investigation and the details provided, atrium medical corporation cannot conclude that the device was at fault.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|