| Model Number |
MMT-722NAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.Customer¿s blood glucose level was unknown.Customer was unable to complete rewind the insulin pump.The customer was advised to discontinued the insulin pump and revert to backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during rewind due to corrode the motor home switch.Unable to perform rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to motor error alarms.
|
| |
|