| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was 117 mg/dl at the time of the incident.The customer was assisted with troubleshooting and the display did not return.Customer stated that the contacts on the battery cap are not missing or damaged.Customer stated that the battery compartment is neither damaged nor corroded.Customer states the spring is neither damaged nor corroded.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test due to blank display.Device received with moisture damage motor, vibrator and battery tube assembly.
|
| |
|